Finally, the clinical sampling process is also very important.For example, in false positives caused by COVID-19 vaccines, before the nucleic acid fragments in the vaccine are completely degraded, if nucleic acid fragments left on the surface of the human body or clothing contaminate the sample, then afterhigh-sensitivity PCR amplificationa false-positive result is very likely to occur.
Positive control materials and weak positive quality control materials must not be added inside the biosafety cabinet; during operation, the size of gloves must be appropriate. Loose gloves are more easily contaminated. If gloves may be contaminated, the gloves should be removed and discarded promptly, and new gloves should be replaced. Frequently changing gloves can effectively reduce various types of contamination.
At present, most laboratories have a large amount of testing work, and cross-contamination is likely to occur during sample addition and extraction. In particular, when samples from key populations (such as quarantined persons) are tested together with samples from the general healthy screening population, false positives are more likely to occur.In this regard, we should focus on the storage and batch testing of population samples and general healthy screening population samples。
During the testing process, we should randomly place 3 normal saline samples in accordance with the requirements of the "Manual for COVID-19 Nucleic Acid Testing in Medical Institutions (Trial Second Edition)" so as to monitor contamination during the experiment.
If the PCR tube caps are not tightly closed or are deformed by heating, the amplification products will evaporate into the PCR instrument and the laboratory, causing a large amount of persistent contamination. In particular, after weak positive quality control materials and weak positive quality control wells leak, it is easier to contaminate the laboratory and produce a large number of false positives.In this regard, we should choose matching PCR tubes and caps that have passed quality inspection。

After sampling, confirm that all PCR tubes are covered and not loose, and then use the machine for amplification; the amplified PCR tubes should be carefully removed and quickly placed in sealed bags for general medical waste disposal. Opening the caps and autoclaving are strictly prohibited.For laboratories prone to false positives, laboratory environmental contamination assessment experiments should be conducted. If the laboratory environment is contaminated, routine laboratory testing should be suspended, and the environment should be cleaned, disinfected, and ventilated.
If the reagents are contaminated by positive quality control materials or positive samples during preparation, the entire plate test will be positive. By replacing the reagents and re-amplifying, it can be determined whether the amplification products are contaminated or the reagents are contaminated. As a preventive measure, we should strictly comply with the laboratory rules and regulations, such as single-flow operation in the laboratory and no mixing of laboratory work clothes and items; at the same time, attention should also be paid to the storage of positive quality control materials and the positive quality control materials mentioned earlier by the author.
After the experiment is completed, we still cannot let our guard down. In addition to publishing the report correctly, to ensure the smooth progress of the next experiment and avoid false positives, laboratory cleaning and 2-hour UV disinfection, regular use of nucleic acid removal agents to treat workbenches and pipettes, and terminal disinfection of the laboratory are also essential.