Immunoassay is a method for analyzing and quantifying a target antibody or antigen based on the principle of highly selective and highly specific recognition and binding between a specific antibody and an antigen or hapten. Enzyme-linked immunosorbent assay (hereinafter referred to as ELISA) is a type of immunoassay and consists of three parts: immune recognition, signal output, and data processing.
Immune recognition involves coating antibodies onto a 96-well plate made of polystyrene, and then using the antibodies to recognize the antigen to be detected (usually disease protein biomarkers, viruses, bacteria, etc.), adsorbing the antigen from the complex sample solution onto the surface of the 96-well plate. Next, antibodies labeled with horseradish peroxidase (HRP), fluorescence, or radioactivity are used to output recognition signals directly or indirectly. Finally, the concentration of the target antigen in the sample to be tested is calculated using the signal intensity, standard sample concentration gradient, and other information.
Depending on the mode of immune recognition and signal output, ELISA can be classified into the double-antibody sandwich method, direct competitive immunoassay, indirect competitive immunoassay, etc. Among these,the double-antibody sandwich methodis the most common.
the double-antibody sandwich method
①: Antibody coating Antibodies are coated onto a 96-well plate. Since the microplate is made of polystyrene, the benzene rings it contains exert an attractive force similar to π-π stacking with the amino acid residues of antibodies, and combined with electrostatic and hydrophobic interactions, antibodies can be adsorbed onto its surface. Then, after washing away unadsorbed antibodies with buffer, a blocking solution containing gelatin or bovine serum albumin (BSA) is added to block the portions of the microplate that have not bound antibodies. The purpose of adding the blocking solution is to prevent other proteins from adsorbing onto the 96-well plate through electrostatic or hydrophobic interactions, which would cause false positive signals and interfere with subsequent experiments.
②: Immune recognition After coating antibodies onto the 96-well plate, the sample to be tested is added and incubated at 37°C for a period of time, usually 1-2 hours. At this point, the antibodies on the microplate specifically recognize and bind to the antigen to be tested. The quality of the antibody is key here; a good antibody can both specifically and efficiently bind the antigen and is not affected by other biological macromolecules, proteins, inorganic salts, and other components in the sample.
③: Plate washing Unbound antigen is washed away, and the recognition antibody corresponding to that antigen is added and incubation is continued at 37°C for 1-2 hours, after which antibodies that have not bound to the antigen are washed away.
④: Enzyme-labeled signal output An enzyme-labeled secondary antibody labeled with horseradish peroxidase (HRP) is added to bind to the standard antibody. After 30 minutes of incubation, unbound secondary antibody is washed away, and a chromogenic substrate is added for color development. The antigen concentration is determined based on the color development results. It is generally believed that the antigen concentration is positively correlated with the luminescence intensity after color development.
Competitive immunoassay
Taking antigen competing for antibody as an example, in the second step of the above sandwich immunoassay procedure, after adding the antigen to be tested, enzyme-labeled antigen is immediately added to compete with the antigen to be tested for recognition of the antibodies on the microplate. The higher the concentration of the antigen to be tested, the fewer enzyme-labeled antibodies can bind to the antibodies on the microplate, and the lower the output signal. In this way, the concentration of the sample to be tested is inversely correlated with the enzyme color signal.