3200威尼斯vip(中国)有限公司

News
NEWS
Home News Industry News
Western blot experiment troubleshooting guide!
2023-10-13


Western Blot is believed to be a routine experiment for many biology research folks, holding a small wish to produce a perfect and beautiful WB image, but the final development may still be an empty, bitter, bloody, and tearful history. The operation process seems simple, but it is interlinked. Even veterans cannot withstand the repeated trials of WB experiments; those who know, know. The following are common problems and solutions encountered during WB experiments, hoping to provide a troubleshooting reference for those currently doing WB experiments!


No bands


1. Low protein expression:

1. Fully understand the protein being studied; if expression is low, appropriately increase the loading amount,20-50μg。

2.whether basal expression or induced expression, nonspecific bands.

3.Use protease inhibitors and phosphatase inhibitors when extracting protein from samples, and set controls to avoid protein degradation.

4.Check whether the primary antibody specificity is suitable for the species of the target protein; read the instructions; it is best to buy validated primary and secondary antibodies.

2. Low protein transfer efficiency:

1. Ponceau S staining can be used to check transfer efficiency,For proteins with small molecular weight, pay attention to using membranes with small pore size.

2.Appropriately adjust and optimize transfer time and current.

3.For low molecular weight expression, choose ultrasensitive luminescent solution.


There are positive bands, but the bands are relatively weak

1.Insufficient antibody staining:Increase antibody concentration and prolong incubation time.

2.Decreased enzyme activity:Mix the enzyme and substrate directly; if no color develops, the enzyme is inactive. Choose an enzyme conjugate within the validity period and with activity.

3.Target protein content in the sample is too low:Increase the sample loading amount.

4.Overwashing of the membrane:Shorten the washing time.

5.HRP Inhibitors:Avoid sodium azide in the solutions and containers used.

6.Decreased antibody activity:Choose antibodies within the validity period; prepare working solution immediately before use and avoid prolonged storage.

7.Overblocking:Reduce the amount of blocking agent or shorten the time; switch to a different type of blocking agent.

8.Exposure time too short:Extend the exposure time.

9.HRP Inhibitors:Avoid sodium azide in the solutions and containers used.


High background

1.Loading amount too high:Appropriately reduce the loading amount.

2.Insufficient membrane washing:Increase wash solution volume and number of washes, the principle can bequick washes with slow incubation.

3. Insufficient blocking:Choose an appropriate blocking reagent,Increase blocking solution incubation time,blocking concentration. Add to the wash solution Tween-20 to reduce cross-reactivity. Casein contained in skim milk powder is a phosphorylated protein that binds to phosphorylation-specific antibodies and easily produces high background. Use BSA instead of milk powder as the blocking agent.

4.Too much antibody:Reduce antibody concentration, use monoclonal antibodies, reduce secondary antibody incubation time

5.Membrane drying during detection:Ensure sufficient reaction solution to avoid membrane drying.

6.Overexposure:Shorten the exposure time.

7.Protein degradation:Use fresh samples, add protease inhibitors, and operate on ice.

8.High background caused by membrane selection:UseLow backgroundNC MembraneReplaceHigh backgroundPVDF Membrane。

9.The protein exists in different spliced forms:Consult database literature to learn about specific primary antibodies for the protein.

Non-specific bands or incorrect protein band size

1.Protein degradation:Addprotease inhibitors, and operate on ice during sample processing.

2.The target protein has multiple isoforms with different molecular weights:Consult literature or databases to learn about the target protein information.

3.The cell line has been passaged too many times:Select a primary cell line with fewer passages for sample preparation.

4.The target protein has multiple modification forms,such as acetylation, methylation, alkylation, phosphorylation, glycosylation, etc. Consult literature or database analysis to obtain modification site information of the protein sequence, and determine whether the protein has different lengths of encodedmRNA。

5.The protein has polymers, such as dimers, tetramers, and multimers:SDS-PAGE Before electrophoresis loadingboil 10 minto enhance protein depolymerization and denaturation。

6.Loading amount too high or antibody concentration too high:Appropriately reduce the loading amount and antibody concentration.

7.Insufficient washing

8.Primary antibody specificity is not high:ReselectOrPreparationGoatHeterophilic antibodies


Issues such as band bending and smearing


Problems encountered

Causes and solutions

Bands appear smile-shaped (︶)

1. Migration too fast or electrophoresis temperature too high: reduce voltage or perform electrophoresis at low temperature.

2. Gel preparation issues, uneven gel cooling, poor cooling in the middle causing fast migration: re-prepare the gel, replace the aged electrophoresis tank with a new one.

Bands appear frown-shaped (︵)

Gel preparation issues, possibly due to improper assembly, especially bubbles between the gel and the bottom of the glass plate, or incomplete polymerization on both sides of the gel.

Bands diffuse or smear

1. Excessive protein loading: reduce the loading amount.

2. Not transferring immediately after electrophoresis: transfer as soon as possible after electrophoresis.

3. Protein degradation: add protease inhibitors.

4. Sample nucleic acid contamination: perform sonication and nuclease treatment, or if the sample contains insoluble matter, dilute it, fully lyse or centrifuge.

Bands skewed

Unbalanced electrodes or off-center sample loading position.

Bands spread to both sides, incomplete

1. Uneven antibody incubation or insufficient binding: increase antibody incubation time, overnight incubation at 4°C is recommended.

2. Antibody concentration too low: use a higher concentration of antibody.

3. Chemiluminescent solution not applied evenly: appropriately increase the amount of chemiluminescent solution.

4. Insufficient reaction time: increase the chemiluminescent color development reaction time.

Hollow bands

 May be caused by substrate depletion due to excessively high protein concentration; diluting the sample concentration is recommended.

Bands with uneven spots

 

1. Bubbles on the membrane: when transferring, slowly apply the PVDF membrane onto the gel, then gently roll with a glass rod to drive out bubbles between the membrane and the gel.

2. Equipment contamination: ensure the equipment is thoroughly cleaned.

3. Insufficient incubation or washing solution: ensure adequate incubation and washing.

4. The blocking solution was not mixed evenly. Ensure thorough mixing, use fresh blocking solution, and avoid particulate insoluble matter.

5. The membrane is contaminated: Laboratory personnel should wear gloves and masks, wear lab coats, and keep the membrane clean. Do not touch the membrane directly with hands.

White reflective bands appear on a dark background

High-concentration HRP in the central area consumes the substrate too quickly, and after the substrate in the central area is exhausted, no luminescence can occur.

1. Too much protein loaded: Reduce the amount of protein loaded.

2. Primary and secondary antibody concentrations are too high: Appropriately increase antibody dilution.

Dumbbell-shaped bands

1. Gel preparation problem; possibly too much force was used when removing the gel, causing the gel to deform and protrude: Prepare a new gel.

2. The protein contains impurities.

Non-uniform background appears, messy

The membrane may have dried out before: Ensure that the membrane remains moist at all times during the experiment.

 


Low protein transfer efficiency

High-molecular-weight proteins

1.Methanol concentration is too high:Reduce the methanol concentration to 10%(V/V) or below.

2.Protein binding time is insufficient:Use wet transfer instead of semi-dry transfer. Prolong the transfer time.

3.Protein gel concentration is too high:Reduce the gel concentration, and add 0.1%SDSto the transfer buffer to promote protein dissolution.

Low-molecular-weight proteins

1.Small proteins bind less to the membrane:Use a membrane with stronger protein-binding capacity. Increase the SDS-PAGE gel concentration.

2.Use a small-pore-size blotting membrane,such as 0.1 μm or 0.2 μm Protran blotting membrane.

3.Residual SDS in the transfer buffer affects protein binding to the membrane:Residual SDS Do not have SDSin it.SDS Before transfer, equilibrate the gel with transfer for at least 15 min 

4.Residual SDS or more.The methanol concentration is too low15%-20%)

5.Protein binding time is insufficient:: increase the methanol concentration (


Lower the voltage and extend the transfer time.


1What membrane should be selected for proteins with very small molecular weight <20KD?) You can choose a membrane with a pore size of 0.22umPVDF membrane or NC Membrane, shorten the transfer time, and in addition, a Tricine-SDS-PAGE system can be used;

2) can also choosePSQ membrane while shortening the transfer time.You can also stack the two membranes together before transferring.


During development, the negative is completely black—why is that?


1) It might be caused by the red light., The film has already been exposed, so it can be handled in complete darkness to see if there is any improvement.

2)Development time is too long,General3-5minutes is enough; for special cases, the color development time needs to be explored.


An inverted image appears on the membrane.

Possible secondary antibody conjugationHRPIf the concentration is too high, the concentration of the secondary antibody can be appropriately reduced.


Should "Ponceau S" or "Coomassie Brilliant Blue" dye be chosen for WB experiments?

Ponceau S The detected protein level is 200 ng The above, and PVDF , compatible with nitrocellulose or nylon membranes. Coomassie Brilliant Blue only with  PVDF Membrane compatible, but can detect 50 ng and above protein. It is recommended to test 50-200 ng For proteins within the range, choose Coomassie Brilliant Blue. However, in  CST, we usually preferPonceau S, because it is fast, easy to use, and gentle yet effective.


Western Blot "quenching"?

1.ContainingHRPAntibodies:In terms of antibodies, this can be addressed by appropriately reducing the antibody concentration and performing the luminescence experiment quickly.“Quenching”'s question

2.Luminescent solution:Can choose a stable luminescenceECLLuminescent solution to solve it.“Quenching”'s problem.


How to choose a protein marker?

1.Based onProteinMarkerMolecular weight range:Choose proteins with as wide a molecular weight range as possible.marker, so that it can be applied to the detection of more proteins. If it has already been determined that the focus is on the detection of large molecular weight proteins or small molecular weight proteins, it is best to choose the corresponding molecular weight rangemarker, in order to get better instructions.

2.Consider using pre-stained proteinMarker:Because pre-stainedMarkerDuring electrophoresis or membrane transfer, it can be directly observed, making it convenient for us to judge the progress of electrophoresis, such as whether abnormalities occur during electrophoresis or whether the target protein is completely separated, etc. It can also determine whether membrane transfer is complete, and after transfer, protein molecular weights can also be directly marked on the membrane.

3.ConsiderMarkerIs the number of bands within the molecular weight range?The internal distribution is uniform, with no extensive blank areas. Whether multiple colors are used to facilitate quick identification of different molecular weight sizes. In addition, the colorfulMarkerIt also adds a touch of brightness to the experimental results. Also, pay attention to choosing strips that are sharp and clear.Marker, so that the molecular weight of the protein can be indicated more precisely.


How should the internal reference be correctly set in WB experiments?

1.Select an appropriate internal reference based on the species of the experimental sample.Mammalian tissues or cells typically useβ-actin,GAPDHetc. For plant samples, you can choosePlant actin、Rubiscoetc.

2.Select an appropriate loading control based on the expression location of the target protein.For example, to detect protein expression in whole cells or the cytoplasm, you can chooseβ-actin,GAPDHandβ-Tubulinas loading controls. To detect proteins in the nucleus, you can choosePCNAandTBPas loading controls. For mitochondrial protein loading controls, chooseCOX IV、VDAC1/Porin. For membrane protein loading controls,Na+/K+-ATPase etc.

3.It is generally best that the molecular weight of the loading control differs from the target protein by5KDor moreto prevent bands from interfering with each other.


Can the primary and secondary antibodies from immunohistochemistry and WB experiments be used interchangeably?

Immunohistochemistry can be used for localization, but it cannot accurately quantify, and sometimes false positives occur, making it difficult to distinguish from background;WBcan specifically detect a certain protein molecule. If the antibody is chosen appropriately, the sensitivity is very high.WBIt can quantify, but cannot localize.

The primary antibodies for the two experiments sometimes cannot be used interchangeably. The company's product instructions generally state which experiments can be performed, and for immunohistochemistry, whether it is suitable for paraffin sections or frozen sections.


Can the same protein sample be used for WB detection of two factors simultaneously?

Yes, sometimes if the antibody has high specificity and high titer, they can be used simultaneously.2Plant one antibody, can detect on one membrane2different proteins.


What should be noted in WB procedures for membrane proteins or cytoplasmic proteins?

If they are membrane proteins and cytoplasmic proteins, the detergent used should be much milder.NeedAddNaFInhibit the activity of phosphorylase, it is recommended to use a commercial mixed inhibitor.


Share to:
Products
Cell Biology
Molecular Biology
Liquid Handling
Bioprocessing
Custom Products
Solutions
Biopharmaceuticals
Cell & Gene Therapy
Vaccines
Diagnostics & Testing
About Us
Company Profile
Corporate Culture